You are viewing the site in preview mode

Skip to main content

Table 2 Mean prediction errors and mean absolute prediction errors of differences between preoperative Estimation of axial length and postoperative axial length

From: Optimizing axial length estimation for intraocular lens power calculation in phacovitrectomy for macula-off retinal detachment

 

AF-OpB (n = 51)

FE-OpB (n = 56)

AF-A scan (n = 56)

S-Algor (n = 56)

A-Algor (n = 56)

Mean prediction errors (mm)

0.73 ± 2.22 (-1.31 to 9.15)

-0.16 ± 0.67 (-3.64 to 1.30)

0.20 ± 0.81 (-2.80 to 4.27)

-0.23 ± 0.64 (-3.64 to 1.30)

-0.13 ± 0.71 (-3.64 to 1.30)

Mean absolute prediction errors (mm)

0.87 ± 2.17 (0 to 9.15)

0.34 ± 0.60 (0 to 3.64)

0.47 ± 0.69 (0.02 to 4.27)

0.31 ± 0.55 (0 to 3.64)

0.36 ± 0.62 (0 to 3.64)

  1. AF-OpB = affected eye’s IOL Master optical biometry, FE-OpB = fellow eye’s IOL Master optical biometry, AF-A scan = affected eye’s A-scan biometry, S-Algor = simple algorithm method, A-Algor = advanced algorithm method